admin 管理员组文章数量: 1086019
2024年12月27日发(作者:layui后台框架)
®
RayBio Mouse Wnt-3a ELISA Kit
Catalog #: ELM-WNT3A
User Manual
Last revised April 1, 2022
Caution:
Extraordinarily useful information enclosed
ISO 13485 Certified
3607 Parkway Lane, Suite 100
Norcross, GA 30092
Tel: 1-888-494-8555 (Toll Free) or 770-729-2992, Fax:770-206-2393
Web: , Email: *******************
1
RayBiotech, Inc.
________________________________________
RayBio
®
Mouse Wnt-3a ELISA Kit Protocol
Table of Contents
Section
I.
II.
III.
IV.
V.
VI.
VII.
VIII.
Introduction
Storage
Reagents
Additional Materials Required
Reagent Preparation
Assay Procedure
Assay Procedure Summary
Calculation of Results
A. Typical Data
B. Sensitivity
C. Spiking & Recovery
D. Linearity
E. Reproducibility
Specificity
Troubleshooting Guide
Page #
3
4
4
4
5
6
7
8
8
8
9
9
10
10
11
IX.
X.
Please read the entire manual carefully before starting your experiment
2
I. INTRODUCTION
The RayBio
®
Mouse Wnt-3a ELISA (Enzyme-Linked Immunosorbent Assay) kit is an in vitro
enzyme-linked immunosorbent assay for the quantitative measurement of mouse Wnt-3a in
serum, plasma and cell culture supernatants. This assay employs an antibody specific for
mouse Wnt-3a coated on a 96-well plate. Standards and samples are pipetted into the wells
and Wnt-3a present in a sample is bound to the wells by the immobilized antibody. The wells
are washed and biotinylated anti-mouse Wnt-3a antibody is added. After washing away
unbound biotinylated antibody, HRP-conjugated streptavidin is pipetted to the wells. The
wells are again washed, a TMB substrate solution is added to the wells and color develops in
proportion to the amount of Wnt-3a bound. The Stop Solution changes the color from blue to
yellow, and the intensity of the color is measured at 450 nm.
Need your results faster? Try RayBiotech's SpeedELISA platform for quantitative detection in
just three hours.
/speedelisa-kits/
Short on sample, or need higher sensitivity? Check out the IQELISA
™
Immuno-PCR platform.
/immuno-pcr/
3
II. STORAGE
The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid
repeated freeze-thaw cycles. The kit may be stored at 4°C for up to 6 months. For extended
storage, it is recommended to store at -80°C. For prepared reagent storage, see table below.
III. REAGENTS
Component
Wnt-3a Microplate (Item A)
Wash Buffer Concentrate
(20X) (Item B)
Standard Protein (Item C)
Detection Antibody Wnt-3a
(Item F)
HRP-Streptavidin
Concentrate (Item G)
TMB One-Step Substrate
Reagent (Item H)
Stop Solution (Item I)
Assay Diluent D (Item K)
Assay Diluent B (Item E)
Size / Description
96 wells (12 strips x 8 wells) coated with anti-
Mouse Wnt-3a.
25 ml of 20X concentrated solution.
2 vials of Mouse Wnt-3a. 1 vial is enough to run
each standard in duplicate.
2 vials of biotinylated anti-Mouse Wnt-3a. Each
vial is enough to assay half the microplate.
200 µl 150X concentrated HRP-conjugated
streptavidin.
12 ml of 3,3,5,5'-tetramethylbenzidine (TMB) in
buffer solution.
8 ml of 0.2 M sulfuric acid.
15 ml of 5X concentrated buffer.
15 ml of 5X concentrated buffer.
Storage / Stability
After Preparation
1 month at 4°C*
1 month at 4°C
1 week at -80°C
5 days at 4°C
Do not store and
reuse.
N/A
N/A
1 month at 4°C
1 month at 4°C
*Return unused wells to the pouch containing desiccant pack, reseal along entire edge.
IV. ADDITIONAL MATERIALS REQUIRED
1.
2.
3.
4.
5.
6.
7.
8.
Microplate reader capable of measuring absorbance at 450 nm.
Precision pipettes to deliver 2 µl to 1 ml volumes.
Adjustable 1-25 ml pipettes for reagent preparation.
100 ml and 1 liter graduated cylinders.
Absorbent paper.
Distilled or deionized water.
Log-log graph paper or computer and software for ELISA data analysis.
Tubes to prepare standard or sample dilutions.
4
V. REAGENT PREPARATION
1.
Bring all reagents and samples to room temperature (18 - 25ºC) before use.
2.
Assay Diluent D (Item K) and Assay Diluent B (Item E) should be diluted 5-fold with
deionized or distilled water before use.
3.
Sample dilution: 1X Assay Diluent D (Item K) should be used for dilution of serum,
plasma, and cell culture supernatant samples. The suggested dilution for normal
serum/plasma is 2 fold.
Note: Levels of Wnt-3a may vary between different samples. Optimal dilution factors for
each sample must be determined by the investigator.
4.
Preparation of standard: Briefly spin a vial of Item C. Add 640 µl 1X Assay Diluent D
(Item K, Assay Diluent D should be diluted 5-fold with deionized or distilled water before
use) into Item C vial to prepare a 62.50 ng/ml standard solution. Dissolve the powder
thoroughly by a gentle mix. Pipette 300 µl 1X Assay Diluent D into each tube. Use the
62.50 ng/ml standard solution to produce a dilution series (shown below). Mix each tube
thoroughly before the next transfer. 1X Assay Diluent D serves as the zero standard (0
ng/ml).
200 µl200 µl200 µl200 µl200 µl200 µl
Std1
Item C +
640 µl
Std2
300 µl
Std3
300 µl
Std4
300 µl
Std5
300 µl
Std6
300 µl
Std7
300 µl
Zero
Standard
300 µl
Diluent
volume
Conc.
62.50
ng/ml
25 ng/ml10 ng/ml4 ng/ml
1.600
ng/ml
0.640
ng/ml
0.256
ng/ml
0
ng/ml
5
5.
If the Wash Concentrate (20X) (Item B) contains visible crystals, warm to room
temperature and mix gently until dissolved. Dilute 20 ml of Wash Buffer Concentrate into
deionized or distilled water to yield 400 ml of 1X Wash Buffer.
6.
Briefly spin the Detection Antibody vial (Item F) before use. Add 100 µl of 1X Assay
Diluent B (Item E) into the vial to prepare a detection antibody concentrate. Pipette up
and down to mix gently (the concentrate can be stored at 4ºC for 5 days). The detection
antibody concentrate should be diluted 80-fold with 1X Assay Diluent B (Item E) and
used in step 5 of Part VI Assay Procedure.
7.
Briefly spin the HRP-Streptavidin concentrate vial (Item G) and pipette up and down to
mix gently before use, as precipitates may form during storage. HRP-Streptavidin
concentrate should be diluted 150-fold with 1X Assay Diluent B (Item E).
For example: Briefly spin the vial (Item G) and pipette up and down to mix gently. Add
80 µl of HRP-Streptavidin concentrate into a tube with 12 ml 1x Assay Diluent B to
prepare a 150-fold diluted HRP-Streptavidin solution (don't store the diluted solution for
next day use). Mix well.
VI. ASSAY PROCEDURE
1.
Bring all reagents and samples to room temperature (18 - 25ºC) before use. It is
recommended that all standards and samples be run at least in duplicate.
2.
Label removable 8-well strips as appropriate for your experiment.
3.
Add 100 µl of each standard (see Reagent Preparation step 3) and sample into
appropriate wells. Cover wells and incubate for 2.5 hours at room temperature with
gentle shaking.
4.
Discard the solution and wash 4 times with 1X Wash Solution. Wash by filling each well
with Wash Buffer (300 µl) using a multi-channel Pipette or autowasher. Complete
removal of liquid at each step is essential to good performance. After the last wash,
remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot
it against clean paper towels.
5.
Add 100 µl of 1X prepared biotinylated antibody (Reagent Preparation step 6) to each
well. Incubate for 1 hour at room temperature with gentle shaking.
6.
Discard the solution. Repeat the wash as in step 4.
7.
Add 100 µl of prepared Streptavidin solution (see Reagent Preparation step 7) to each
6
well. Incubate for 45 minutes at room temperature with gentle shaking.
8.
Discard the solution. Repeat the wash as in step 4.
9.
Add 100 µl of TMB One-Step Substrate Reagent (Item H) to each well. Incubate for 30
minutes at room temperature in the dark with gentle shaking.
10.
Add 50 µl of Stop Solution (Item I) to each well. Read at 450 nm immediately.
VII. ASSAY PROCEDURE SUMMARY
1.
Prepare all reagents, samples and standards as instructed.
2.
Add 100 µl standard or sample to each well. Incubate 2.5 hours at room temperature.
3.
Add 100 µl prepared biotin antibody to each well. Incubate 1 hour at room temperature.
4.
Add 100 µl prepared Streptavidin solution. Incubate 45 minutes at room temperature.
5.
Add 100 µl TMB One-Step Substrate Reagent to each well. Incubate 30 minutes at
room temperature.
6.
Add 50 µl Stop Solution to each well. Read at 450 nm immediately.
7
VIII. CALCULATION OF RESULTS
Calculate the mean absorbance for each set of duplicate standards, controls and samples,
and subtract the average zero standard optical density. Plot the standard curve on log-log
graph paper or using Sigma plot software, with standard concentration on the x-axis and
absorbance on the y-axis. Draw the best-fit straight line through the standard points.
A. TYPICAL DATA
These standard curves are for demonstration only. A standard curve must be run with each
assay.
B. SENSITIVITY
The minimum detectable dose of Mouse Wnt-3a was determined to be 0.21 ng/ml.
Minimum detectable dose is defined as the analyte concentration resulting in an absorbance
that is 2 standard deviations higher than that of the blank (diluent buffer).
8
C. SPIKING & RECOVERY
Recovery was determined by spiking various levels of Mouse Wnt-3a into the sample types
listed below. Mean recoveries are as follows:
D. LINEARITY
9
E. REPRODUCIBILITY
Intra-Assay CV%: <10%
Inter-Assay CV%: <12%
IX. SPECIFICITY
This ELISA antibody pair detects mouse Wnt-3a. Other species not determined.
10
X. TROUBLESHOOTING GUIDE
Problem
Poor standard
curve
Cause
Inaccurate pipetting
Improper standard
dilution
Solution
Check pipettes
Briefly centrifuge Item C and dissolve
the powder thoroughly by gently mixing
Low signal
Improper preparation of
standard and/or
biotinylated antibody
Too brief incubation
times
Inadequate reagent
volumes or improper
dilution
Briefly spin down vials before opening.
Dissolve the powder thoroughly.
Ensure sufficient incubation time. Assay
procedure step 3 may be done
overnight at 4°C with gentle shaking
(note: may increase overall signals
including background).
Check pipettes and ensure correct
preparation
Large CV
Inaccurate pipetting
Air bubbles in wells
Check pipettes
Remove bubbles in wells
High
background
Plate is insufficiently
washed
Contaminated wash
buffer
Review the manual for proper wash. If
using a plate washer, ensure that all
ports are unobstructed.
Make fresh wash buffer
Low sensitivity
Improper storage of the
ELISA kit
Stop solution
Store your standard at <-70ºC after
reconstitution, others at 4ºC. Keep
substrate solution protected from light.
Add stop solution to each well before
reading plate
11
RayBio
®
ELISA Kits
Over 4,000 ELISA kits available, visit /ELISA-Kits for details.
This product is for research use only.
©2020 RayBiotech, Inc
12